Specifications
Insulation Resistance
- Test voltage: 250V / 500V / 1000V.
- Insulation Resistance: 0.2M~2GΩ / 0.2M~4GΩ / 0.2M~8GΩ.
- Accuracy: 0.2MΩ~4GΩ: ±3%rdg. 4GΩ~8GΩ: ±5%rdg.
- Nominal voltage rated: 1mA.
- Short circuit current: 1.2mA.
- PI (Polarization Index).
- DAR (Dielectric Absorption Ratio).
Continuity
- Range: 0.01~100Ω / 100~300Ω / 300~1999Ω.
- Accuracy: ±1%rdg / ±1.5%rdg / ±2%rdg.
- Resolution: 2 counts.
- Auto-Null Up to 5Ω.
- Buzzer Threshold Value Up to 3Ω.
Voltmeter (Autoranging)
- Range: 0~700VAC / 0~950VDC.
- Resolution: 1V.
- Accuracy: ±1.5% F.S.
KΩ Test
- Range: 1~400kΩ.
- Short circuit test current: ≥1.3mA.
Diode Test
- Max test current: 1.5 mA.
- Resolution: 0.1V.
- Measurement voltage: 0-4.5V.
- Accuracy: ±3% rdg.
MOVs & GAS Voltage Result @ 1mA.
- Test Voltage: 5 ~ 1020V.
- Accuracy: ±3%rdg±2dgt.
General
- Fuse: 500mA 250V (5 × 20mm) HBC. Fast Blow.
- Display: 2 lines × 16 characters LCD.
- Operating / Storage temp.: 0°C~40°C / -10°C~50°C.
- Dimensions: 205(L) × 90(W) × 55(D)mm.
- Weight (battery included): Approx. 1460g.
- Power source: 1.5V (AA) × 6 Alkaline batteries.
- Safety Standard: EN 61010-1 CAT III 600V. EN 61326-1.
Accessories Instruction manual, Test leads, Carry Case, Batteries.