Specifications
Determination of the Phase Presence
- Nominal voltage for phase presence: 100Vac ~ 450Vac.
- Frequency Range: From 10Hz to 400Hz.
Determination of the Phases Rotary Field Direction
- Direction: From 100 to 450Vac.
- Frequency Range: From 2Hz to 400Hz.
Protection
- Over Load: 550V(between all terminals).
- Over Voltage: Class III - 450V towards ground.
Earth Leakage
- Current Settings: 999mAac / 50Hz-60Hz.
- Current Selection: Knob.
- Phase Start Selection (Referenced to Earth): 0° and 180°.
- Over-Temperature Protection(Referenced to Earth): Yes.
- Phase Polarity Trip Indicator(Referenced to Earth): Yes.
- Operating Voltage(L-E): 220Vac to 317Vac.
- Timer Resolution: 1ms (Max. Time = 99.99s).
- Timer Accuracy: 1ms ± 4ms.
- Current Accuracy: ±3% ± 4mA.
- Current Resolution: 1mA.
- Voltmeter Accuracy (50Hz):
- 50-350 Vac = ±3% ± 2V.
- 350-550Vac = ±7% ± 3V.
- Voltmeter Resolution: 1V.
- Maximum Current Specified at 317Vac / 50Hz.
General
- Operating temperature: 0°C to 40°C.
- Storage temperature: -10°C to 50°C.
- Dimensions: 330(L) × 260(W) × 160(D)mm.
- Weight (battery included): Approx. 3850g.
- Batteries: 1.5V "C" × 8 Alkaline batteries. Bat OK Led = Vbat >9V.
- Safety Standard: EN 61010-1 CAT III 550V. EN 61326-1.